Copan Wasp: Calculator/data Processing Module, For Clinical Use

Recall Summary Copan WASP is recalling WASPLab automated microbiology devices with specific software versions (2.33.1, 2.33.3, and 2.34.2) because a software issue may prevent users from correcting or overwriting test results. The recall affects devices distributed nationwide across all U.S. states and the District of Columbia. This is classified as a serious recall (Class I).

Recalling firm
Copan WASP
Product
Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2 Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with microbiological specimens derived by the human body for microbiological investigations. WebApp software is the user interface of WASPLab device.
Reason
Due to a software issue within the WebApp that potentially prevents the user to overwrite the results.
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-16
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the state of Arizona.