Tobii Dynavox Llc: System, Communication, Powered
Recall Summary Tobii Dynavox is recalling the TD Navio Maxi Speech Generating Device (Model 1000799) sold nationwide in the US because the battery mounting tabs can become damaged, allowing the battery to shift inside the device and potentially cause an internal short circuit that may result in localized heating and gas venting. This recall affects users across all 50 states and Washington, DC.
- Recalling firm
- Tobii Dynavox Llc
- Product
- TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
- Reason
- Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-07-08
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- US Nationwide distribution.