All Lifts With Weigh Scale: Tenor

Recalling firm
Arjo New Zealand Limited
Product
ALL LIFTS WITH WEIGH SCALE: Tenor. [Device]. Affected: Tenor lifts manufactured between 06-Mar-2025 and 28-Nov-2025, and spare part actuators (TEN.107 and RKT.560).
Reason
During final product testing at the manufacturing site, an anomaly in the behaviour of the lifting actuator was detected. The subsequent testing confirmed that an internal component of the actuators in the certain production batch may not meet the required mechanical strength. This could result, in a sudden loss of the actuator's ability to hold the load, leading to rapid and uncontrolled downward movement of the lifting arm., Since this internal actuator's malfunction cannot be identified through standard pre-use check and only becomes evident at the moment of failure, the issue is not...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-10
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory