Rsch-uchida Transjugular Liver Access Sets, Reference Part Numbers: Rups-100; The Rsch-uchida Transjugular Liver Access Set Is Intended For Transjugular Liver Access In Diagnostic And Interventional

Recall Summary Cook Medical is recalling Rsch-Uchida Transjugular Liver Access Sets (model RUPS-100), which are medical devices used to access the liver during diagnostic and treatment procedures. Some units from affected production lots may have been cut to incorrect lengths, improperly trimmed, or inadequately inspected during manufacturing. This Class II recall affects medical facilities that received these products.

Recalling firm
Cook Incorporated
Product
Rsch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The Rsch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.
Reason
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
Classification
Class II
Category
consumer product recall
Country
United States
Agency
FDA-DEVICES
Recall date
2025-11-26
Status
Open, Classified

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/74107241-ff00-4b95-918b-be8c85b58d35