Hologic - Horizon W X-ray Bone Dxa Systems Wi
- Recalling firm
- Hologic (Australia and New Zealand) Pty Ltd
- Product
- Hologic - Horizon W X-Ray Bone DXA Systems WI. [Device]. Model: Part number: Horizon-W. Affected: Lot number: 308911M
- Reason
- The Horizon WI systems manufactured from 11 December 2024 to 16 January 2025 were shipped without the validated VGA cables. Some installations were completed using VGA cables that may not meet Hologic's applicable performance and safety standards for this system. The impacted units may present a potential risk of elevated Electromagnetic Interference (EMI) to users and/or patients near the affected Horizon DXA systems. The impact can range from user dissatisfaction to delay in treatment and presents an elevated risk of interference to the population of individuals with active, implantable...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-10-27
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory