Lutronic - Clarity II & Clarity II Lpa

Recalling firm
Ebos Medical Devices NZ limited
Product
Lutronic - CLARITY II & CLARITY II LPA. [Device]. Model: LARITY II & CLARITY II LPA. Affected: Serial number - C3025C065
Reason
The manufacturer has identified a potential safety-related issue may occur when operating the CLARITY II ICD system under high-fluence conditions (= 90-100 J/cm2, 1064nm) with cryogen cooling. Under these specific operating conditions, spark or loud noise (pop sound) may be observed during treatment and may increase the likelihood of localized minor burn/blister on the skin.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-18
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE