Haemofilter

Recalling firm
Vantive Limited
Product
Haemofilter. [Device]. Model: Prismaflex ST100 and Prismaflex ST150 - Product code 107636, 107640; Prismaflex TPE2000 - Product code 107144; Oxiri - Product code 97003; (Oxiris). Affected: Prismaflex ST100 - All lots including and manufactured after 24F0069. Prismaflex ST150 - All lots including and manufactured after 24E0059; All lots including and manufactured after 24E0067CB; All lots including and manufactured after 24F0087. Also covers: Meyzieu TPE sets; oXiris Set - EU MDR; oXiris Set - EUMDR - Vantive
Reason
The manufacturer is notifying Prismaflex Control Unit users of a potential risk of the Prismaflex Set deaeration chamber becoming dislodged during use. This issue does not affect PrisMax Control Units and may only occur when Prismaflex Sets are used with the Prismaflex Control Unit, as the two systems have different deaeration chamber holder designs.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-26
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/725dabe1-a299-410c-8a23-dcbad2f1749d