Hemodynamic Application (mdr)

Recalling firm
Philips New Zealand Commercial Limited
Product
Hemodynamic Application (MDR). [Device]. Model: Philips Interventional Hemodynamic System; Affected systems are Philips Interventional Hemodynamic Systems with Philips Hemodynamic Application software version 1.4.1 or lower. Also covers: IntelliVue Patient Monitors
Reason
The Philips Interventional Hemodynamic System automatically deletes the oldest case data when storage is full, but it does not remove export files, snapshots, or generated reports. These accumulate over time and can eventually fill the Hemo Host PC disk. When the disk is full, new cases cannot be opened - the application launches in review mode only. Live signals from the IntelliVue X3 will still display, but users cannot interact with the data. If the disk becomes full during an active case, data will continue to display and functionality remains, but the case will not be saved and cannot...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-12
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory