Fresenius Kabi Usa, Llc: Lvp Software Of The Ivenix Infusion System (iis); Product Code Lvp-sw-0005

Recall Summary Fresenius Kabi is recalling the LVP software (Product Code LVP-SW-0005) used in the Ivenix Infusion System pump due to a software problem that can cause the pump to display an incorrect battery level and shut down unexpectedly, requiring it to be plugged into AC power to restart. This recall affects infusion pumps distributed nationwide across multiple states. The company classified this as a Class I recall, the most serious category.

Recalling firm
Fresenius Kabi USA, LLC
Product
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
Reason
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Classification
Class I
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-05-06
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.