Abiomed, Inc.: Introducer, Catheter

Recall Summary Abiomed is recalling Impella CP Introducer Kits (14 Fr size, 13 and 25 cm lengths) used with heart pump devices worldwide due to potential leakage from the introducer sheath under the cap and along hub seams. The leakage cannot be seen by users, and these introducers are included in every pump set. This is classified as a Class II recall affecting patients and facilities in the US and numerous countries internationally.

Recalling firm
Abiomed, Inc.
Product
ABIOMED Impella CP Introducer Kit 14 Fr, 13 and 25 cm, Pump Set Code 1000550, containing Kit, 14Fr Introducer, 13cm&25cm, Sterile, Introducer Product Code 0052-3046. heart pump and accessories
Reason
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-11
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT),...