Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All Versions: R1.x, R2.x, R3.x

Recall Summary Philips is recalling Azurion 5 M12 medical imaging systems (model numbers 722227 and 722231) with software versions R1.X, R2.X, or R3.X because the examination table may move unexpectedly when the Reset Geometry button is pressed, even if the table lock is engaged. This recall affects healthcare facilities using these systems.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
Product
Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X
Reason
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Classification
Class II
Category
consumer product recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-01-26
Status
Open, Classified

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/6fed09bd-44c4-4098-bffc-89b84ddf92f5