Medline Industries, Lp: Circuit, Breathing (w Connector, Adaptor, Y Piece)

Recall Summary Medline Convenience kits with model numbers DYNJ905503D, DYNJ905503F, and DYNJ905503G are being recalled because they contain Swan-Ganz catheters manufactured by Becton Dickinson that have a defective blue hub on the proximal injectate lumen, which can leak or break. The defect may result in leakage, medication loss, blood loss, or infection. These kits were distributed nationwide.

Recalling firm
Medline Industries, LP
Product
Medline Convenience kits containing recalled Swan-Ganz Catheters CV ANESTHESIA - ROOM SET UP DYNJ905503D DYNJ905503F DYNJ905503G
Reason
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-28
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.