Philips Medical Systems Nederland B.v.: Interventional Fluoroscopic X-ray System

Recall Summary Philips is recalling certain Azurion 7 B20 X-ray systems (model numbers 722068 and 722226) worldwide because unapproved solid-state drives may have been installed in the equipment's computers. These drives could cause the system to lose imaging functionality, display a blank screen, or fail to save or retrieve images. This is a Class II recall affecting facilities in the US and numerous countries globally.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Azurion 7 B20. Model Numbers: 722068, 722226.
Reason
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-10-06
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands,...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/6b1e014c-2807-4acd-b263-25fa2991a8f8