Philips Medical Systems Nederland B.v.: Interventional Fluoroscopic X-ray System
Recall Summary Philips is recalling certain Azurion 7 B20 X-ray systems (model numbers 722068 and 722226) worldwide because unapproved solid-state drives may have been installed in the equipment's computers. These drives could cause the system to lose imaging functionality, display a blank screen, or fail to save or retrieve images. This is a Class II recall affecting facilities in the US and numerous countries globally.
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Philips Azurion 7 B20. Model Numbers: 722068, 722226.
- Reason
- Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-10-06
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands,...