Pride Mobility Products Corp.: Wheelchair, Powered
Recall Summary Pride Mobility is recalling Quantum Rehab Q6 Edge 3 Stretto powered wheelchairs with software version 2.0.11 due to a problem with reminder notifications. When users select "Remind Me Later" for reminders, pop-up messages may appear while the wheelchair is operating and cause it to suddenly stop or crash, potentially resulting in falls or injuries. The recall affects wheelchairs distributed nationwide across the US, Canada, and Australia.
- Recalling firm
- Pride Mobility Products Corp.
- Product
- Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
- Reason
- If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-06-25
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and...
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/6a87cc8f-9236-4f26-8fab-4841e1dee77e