Philips North America: System, Nuclear Magnetic Resonance Imaging

Recall Summary Philips is recalling certain Achieva 1.5T MRI machines (models 781196 and 781296) distributed worldwide because software settings can cause incorrect stiffness measurements when using MR Elastography, a specialized imaging technique. The error occurs when specific image reconstruction parameters are combined with Resoundant's algorithm, resulting in voxel size settings that display too small. This recall affects hospitals and imaging centers in the United States and numerous countries internationally.

Recalling firm
Philips North America
Product
Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-14
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar,...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/66dab3bf-89ef-45fd-93ab-32343696fd23