Bd Switzerland Sarl: Stopcock, I.v. Set

BD Connecta BD Luer-Lok 360 stopcocks, which are IV devices used to deliver fluids, medications, and nutritional products, are being recalled because the connection port may over-thread or continue spinning instead of securing tightly when used with plastic syringes. This threading issue has caused some clinical users to be concerned about connection stability. The recall affects devices distributed across most U.S. states and Japan.

Recalling firm
BD SWITZERLAND SARL
Product
BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.
Reason
Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-21
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, FL, GA, IL, IN, IA, KS, ...
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Japan