Horus Tmc
- Recalling firm
- Evolutis New Zealand Limited
- Product
- Horus TMC. [Device]. Model: HORUS TMC cup. Affected: LOT numbers: 2010965, 2016014, 2017024, 2022633, 2021733
- Reason
- The device size is engraved on the back of the insert. A deviation has been observed in the depth of this engraving, which poses a potential risk of the thickness being less than the approved thickness., The potential consequence is premature wear of the insert. All new batches manufactured since May 2026 comply with the requirements (no size engraving). Affected product that has not been implanted has been destroyed and replaced with new the new batch references so you will not receive any affected batches moving forward.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-06-29
- Status
- Ongoing
- Distribution
- New Zealand - Healthcare Professional
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/65013b60-75e3-442a-8efd-d70b5260c3d5