Alaris System Infusion Pumps_cfn San Diego
- Recalling firm
- Becton Dickinson Ltd
- Product
- Alaris System Infusion Pumps_CFN San Diego. [Device]. Model: BD Alaris™ Point of Care Unit (PCU). Affected: Software - 12.1.1
- Reason
- The manufacturer has identified that the device labelling does not indicate which Transport Layer Security (TLS) versions must be enabled on the hospital network to ensure secure Wi-Fi communication with the AlarisTM and BD Alaris™ Point-of-Care Unit (PCU) when using enterprise-level Wi-Fi security. Without clear labelling, hospitals may unknowingly disable TLS 1.0, resulting in network connectivity failure and memory leaks which could lead to potential System Error., Alaris™ and BD Alaris™ PCUs (v12.1.1 and prior) are not compatible with TLS security frameworks beyond TLS 1.0. If TLS 1.0...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-09-24
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE