Synergy Grafts
- Recalling firm
- Getinge Australia Pty Ltd (NZ)
- Product
- Synergy Grafts. [Device]. Model: Intergard Synergy Knitted. Affected: Lot: 1690504264 - The affected products listed are manufactured from April to July 2026
- Reason
- A defect was identified within the body of Intergard Knitted straight vascular grafts from specific textile batches. The defect in question is small loops of yarn within the lumen of the grafts, which are not visible from the external side of the grafts., The presence of one or more yarn loops within the lumen of an Intergard Standard or Intergard Synergy Knitted vascular graft is associated with a low likelihood of causing clinically significant adverse consequences. The defect is not expected to be readily detectable by the surgeon or perioperative staff before or during implantation;...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-08-31
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/610a8c3a-2df2-477c-a1c6-e9de499babd0