Civco Medical Instruments Co. Inc.: System, Imaging, Pulsed Echo, Ultrasonic

Recalling firm
Civco Medical Instruments Co. Inc.
Product
eTRAX Needle Sensor - 16G(for Aurora Trackers), Part Number 667-158
Reason
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-03-02
Status
Open, Classified
Areas affected
OH, PA, WA
Distribution
US States: OH, PA, WA. China.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/5f6e8808-4339-49b1-96b5-d63a6469202c