Eye Tray

Recall Summary Medline is recalling two convenience kits (BASIC EYE and TUBAL LIGATION CDS) distributed nationwide that contain a recalled adhesive product made by Ferndale Laboratories. The adhesive's tubing may crack during use, making the product unable to function properly. The recall affects healthcare facilities and providers across all U.S. states and territories.

Product
Medline convenience kits: BASIC EYE DYNJ905154I TUBAL LIGATION CDS CDS760053K
Reason
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-27
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution. OUS distribution pending.