Recalling firm
Beckman Coulter NZ Ltd
Product
Access Total T4 Calibrators. [Device]. Affected: Lots - 538729, 538815, 538898, 539107, 539243
Reason
The manufacturer has determined that the Access Total T4 Calibrator may produce erroneously low results on UniCel DxI 600 and UniCel DxI 800 Access Immunoassay Analyzers. Internal testing has determined that calibration curves generated using the Access Total T4 Calibrator may produce quality control and patient results approximately 23% lower than expected on UniCel DxI 600 and UniCel DxI 800 Access Immunoassay Analyzers. Falsely decreased Total T4 results reported to the physician may result in patient misdiagnosis and inappropriate treatment.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-19
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE