Generic Name: Washer-disinfector For Flexible Endoscopes — Trade Name: Endoscope Washer-disinfector Espal+

Recalling firm
Name of Manufacturer/Distributor: Toyoei Kogyo Co., Ltd. — Address of Manufacturer/Distributor: 1-369 Shinma-hira, Kawada, Shinshiro City, Aichi Prefecture — License Type: Type II Medical Device...
Product
Lot: ESPP00005、ESPP00006、ESPP00007、ESPP00008、ESPP00009、ESPP00010、ESPP00012、ESPP00014、ESPP00015、 ESPP00016、ESPP00017、ESPP00019、ESPP00020、ESPP00021、ESPP00022、ESPP00023、ESPP00024、ESPP00025、 ESPP00026、ESPP00027、ESPP00028、ESPP00029、ESPP00030、ESPP00032、ESPP00033、ESPP00034、ESPP00035、 ESPP00036、ESPP00038、ESPP00040、ESPP00043、ESPP00045、ESPP00046、ESPP00047、ESPP00048、ESPP00049、 ESPP00050、ESPP00051、ESPP00052、ESPP00053、ESPP00054、ESPP00055、ESPP00057、ESPP00058、ESPP00059、 ESPP00060、ESPP00061、ESPP00063、ESPP00064、ESPP00066、ESPP00067、ESPP00068、ESPP00069、ESPP00070、...
Reason
This product was originally designed to display an error message when no disinfectant solution (undiluted) is drawn in; however, since it lacked a detection function to identify instances where the drawn-in volume was below the specified amount, we have decided to modify the product to ensure that an error is reliably displayed.
Classification
Class II
Category
medical device recall
Country
Japan
Agency
PMDA-DEVICE
Status
Ongoing

Official recall notice from PMDA-DEVICE

Recall details on Nibble Recalls: https://trynibble.app/recalls/599800d5-c6be-4d27-b6b4-1c1913f42b30