Reflexion Medical, Inc.: Fludeoxyglucose F18-guided Radiation Therapy System

Recalling firm
Reflexion Medical, Inc.
Product
RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
Reason
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-02-09
Status
Open, Classified
Areas affected
CA, CT, LA, NJ, OH, OR, PA, TX
Distribution
US: CA, TX, PA, CT, NJ, OR, LA, OH

Official recall notice from FDA-DEVICES