Ge Medical Systems, Llc: Incubator, Neonatal

Recall Summary GE HealthCare is recalling certain Giraffe OmniBed neonatal incubators (Model Numbers 6600-1056-400 and 6600-1461-500) because fasteners securing the canopy soffit under the control panel can loosen or come off during transport or impact, potentially falling into the patient compartment or causing the canopy to detach. The devices are distributed nationwide across the United States. This could potentially result in patient injury.

Recalling firm
GE Medical Systems, LLC
Product
Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500
Reason
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-01
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/558302ab-40d3-4496-a940-38ba593d8553