Corpuls 3 Defib Including Accessories And Upgrades
- Recalling firm
- Universal Specialities Ltd trading as USL Medical
- Product
- Corpuls 3 Defib including Accessories and Upgrades. [Device]. Model: corpuls3 SLIM/SLIM TOUCH Defibrillator Module. Affected: Product codes: 04301, 04302. All serial numbers
- Reason
- The manufacturer has identified a potential problem associated with the magnetic charging interface (MagCode) of the corpuls3-SLIM/SLIM Touch Modules., Increased contact resistance may develop within the magnetic charging interface of the corpuls3-SLIM/SLIM Touch. This may occur due to:, · material combinations with differing thermal expansion characteristics;, · electrochemical effects such as oxidation or corrosion arising from different material potentials, · wear, contamination, or aging of the contact surfaces., As a result, the following effects may occur:, · localised heat...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-07-27
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/525304a9-6063-4e5a-b49e-d5732c0b5151