Omni Legend System Images

Recalling firm
GE Healthcare Limited
Product
Omni Legend System Images. [Device]. Model: Omni Legend PET/CT, all configurations using Software version SP5,SP6,SP7,SP8
Reason
The manufacturer has become aware of a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. In some cases, the artifact can mimic pathology and has the potential to lead to misdiagnosis if undetected. (Note: Standalone CT image acquisition is not impacted by this issue.)
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-12
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/520f901b-3d22-4c37-be6d-c040fe7270e7