Rida Setup 3.01.05

Recalling firm
AusDiagnostics NZ Ltd
Product
RIDA Setup 3.01.05. [Device]. Model: RIDA Xplore System (96521) and HighPlex System (91501)
Reason
A software defect was identified in RIDA® Setup v3.01.05, affecting the RIDA®Xplore System (96521) and HighPlex System (91501) The defect occurs only when all the following conditions are met:, · Samples are entered manually;, · The POSCON option is selected; and, · The positive control is placed in any position other than the last position in the run., Under these conditions, the software repositions a sample without updating the corresponding sample identifier, resulting in a mismatch between the physical sample position and the assigned sample label. This creates the potential for...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-04
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE