Olympus - Ligating Device, Single-use (japan)

Recalling firm
Olympus (NZ) Ltd
Product
Olympus - Ligating device, single-use (Japan). [Device]. Model: SINGLE USE LIGATING DEVICE POLYLOOP (HX-400U-30). Affected: Multiple Lot numbers
Reason
The manufacturer previously communicated a Product Correction (Medsafe Ref #36740) related to additional clarifications in the Instructions for Use (IFU) for proper release of the Single Use Ligating Device (HX-400U-30). Alternatively, customers may have received a Notification of IFU Change with recently purchased products. The action was taken based on complaints from customers indicating that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally anchored in place around patient anatomy. The manufacturer is introducing a...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-09-21
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE