Intelerad-inteleradpacs

Recalling firm
Clinical & Regulatory Services Limited
Product
Intelerad-InteleradPACS. [Device]. Model: IntelePACS - IntelePACS Image Lifecycle Manager (ILM)
Reason
The manufaturer has become aware of a problem affecting certain versions of the IntelePACS Image Lifecycle Manager (ILM). ILM allows administrators to define rules that select studies as deletion candidates based on criteria such as organisation, modality, and study age., When a rule condition includes multiple values, the system may use partial matching rather than requiring an exact match. As an example, if the rule is intended to match organisations ABC or DEF for deletion, organisation ABCD may also be identified for deletion. This may result in the unintended deletion of studies from...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-02
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE