Atellica Ch And Atellica Ci Analysers

Recalling firm
Siemens Healthcare (NZ) Ltd
Product
Atellica CH and Atellica CI Analysers. [Device]. Model: Atellica CH HDL Cholesterol (HDLC) and Atellica CH Total Protein_2 (TP_2). Affected: Material Number - 11537213 and 11537228 All Lots
Reason
The manufacturer has confirmed via complaints the potential for reagent carryover on the Atellica CH and Atellica CI Analysers. If carryover occurs, it may affect patient samples, quality control (QC), and/or calibrator results., For the Atellica CH Analyzer:, · Carryover from Atellica CH HDL Cholesterol (HDLC) may cause falsely elevated Atellica CH Enzymatic Haemoglobin A1c (A1c_E, A1c_H) results., For the Atellica CI Analyzer:, · Carryover from Atellica CH HDL Cholesterol (HDLC) may cause falsely elevated Atellica CH Enzymatic Haemoglobin A1c (A1c_E) results., · Carryover from Atellica...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-10-12
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/48f64e2d-9300-4960-b87e-c423236a96b6