Azurion
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- Azurion. [Device]. Model: All Philips Azurion systems (R1.X, R2.X, R3.X, R4.0) are affected.; All Philips Azurion systems (R1.X, R2.X, R3.X, R4.0) are affected.. Also covers: Angiographic X-Ray
- Reason
- The manufacturer has identified a potential safety issue when the Reset Geometry button is configured to overwrite the table lock, while the Philips system is being used with a stand-alone third-party device or a stationary accessory not synchronized with the table (e.g., a floor-mounted device, a head stabilizer, a robotic arm with a biopsy needle). If the Reset Geometry button is then (accidentally) pressed, unintended table movement may cause the patient to move relative to the stand-alone third-party device or stationary accessory, posing a risk to the patient (e.g., if a needle is...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-03-02
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE