Cervical/endometrial Applicator
Recall Summary Nucletron BV is recalling a cervical/endometrial applicator used in hospitals and pharmacies in New Zealand due to a need to issue updated guidance on its use. This is classified as a Class I recall, which is the most serious type. The manufacturer will be providing advice to healthcare facilities about how to properly use this medical device.
- Recalling firm
- Elekta Pty Limited - New Zealand Branch
- Product
- cervical/endometrial Applicator. [Device]. Model: Brachytherapy applicator, remote afterloading, Cervical/Endometrial
- Reason
- The manufacturer are informing customers of a safety concern regarding the detachment of the tip of the Intrauterine tube (IU tube) with a 4mm diameter. Inadequate application of the epoxy adhesive between the tip and the inner and end surfaces of the fibre tube may contribute to tip detachment., If the tip detaches during a procedure and is retained in the body there is risk of reduced fertility, inflammation and pain.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-04-01
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/451639bf-79cb-4a53-a53b-f61d339b0a98