Anaesthesia Work Stations

Recalling firm
GE Healthcare Limited
Product
Anaesthesia Work Stations. [Device]. Model: Carestation 850 and 850c Anesthesia Systems. Affected: Affected Replaceable Unit part: 5830520-S - SELECTOR VALVE ASSEMBLY
Reason
The manufacturer has become aware that certain Carestation 850 and 850c Anesthesia Systems, can deliver less anesthetic agent than the user selected concentration setting. Reduced delivery of anesthetic agent can only occur when two vaporizers (either Sevoflurane or Isoflurane) are installed on the system, and the left mounted vaporizer is selected for anesthetic agent administration., This issue can be detected during preoperative checkout or while in clinical use., In the unlikely case this issue occurs during clinical use and the audible alarm and visual alarm messages are not noticed,...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-02
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE