Icu Medical, Inc.: Pump, Infusion

Recall Summary ICU Medical is recalling the Plum Duo Precision IV Pump (model 40002-0403) due to a software issue that may prevent the pump from alerting clinicians when a soft limit is violated before delivery begins. This could result in patients receiving incorrect medication doses—either too much or too little. The pump has been distributed nationwide in the US and Canada.

Recalling firm
ICU Medical, Inc.
Product
Plum Duo Precision IV Pump, 40002-0403
Reason
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-29
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.

Official recall notice from FDA-DEVICES