Carescape Patient Monitor, Central Unit
- Recalling firm
- GE Healthcare Limited
- Product
- Carescape Patient monitor, central unit. [Device]. Model: Product code: STU CSCS V3 MAS800 Desktop, Product code: STV CSCS V3 MAI800 Integrated. Affected: Part number: 5867474-04, 5867474-013, 5867474-03
- Reason
- The manufacturer has become aware that the Carescape Central Station (CSCS) V3.0.x may not resume patient monitoring for at least 3 minutes following a system restart. This situation can arise only when the Allow list database is corrupted message is displayed in the title bar of the CSCS prior to a system restart. In the unlikely event that this situation occurs, it can lead to a potential delay in recognition of clinically significant events.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-08-21
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/42bfb0d6-4da7-4379-bd8c-da7cac7460ad