Accuray Incorporated: Cyberknife Treatment Deliver System Equipped With The Xchange Robotic Collimator Changer; Linear Accelerator

Cyberknife Treatment Delivery System Recall Summary Accuray is recalling Cyberknife Treatment Delivery Systems equipped with the Xchange Robotic Collimator Changer due to a safety issue where a collimator may not fully dock properly, and if operators use the in-room control pendant after this happens, the collimator can fall from its housing. This recall affects medical facilities worldwide, including all U.S. states, as well as facilities in China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, and Ukraine.

Recalling firm
Accuray Incorporated
Product
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Reason
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-06-16
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.

Official recall notice from U.S. Food and Drug Administration