Atellica Ch Enzymatic Hemoglobin A1c (a1c_e)

Recalling firm
Siemens Healthcare (NZ) Ltd
Product
Atellica CH Enzymatic Hemoglobin A1c (A1c_E). [Device]. Affected: All in-date lots
Reason
The manufacturer has confirmed, through internal investigation of customer complaints, the potential for intermittent well-to-well bias affecting the A1c_E/A1c_H assay when used on the Atellica CH and Atellica CI systems. This observed bias does not affect every A1c_E reagent lot or reagent pack. However, affected wells cannot currently be identified prospectively, therefore all in-date lots are considered potentially impacted until further notice. Under certain circumstances, this problem may result in:, 1. Falsely depressed patient results when lot calibration is performed using an...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-06-11
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE