Trios 6 System (l1p-2)

TRIOS 6 System Recall Summary The TRIOS 6 System (L1P-2) is being recalled by 3Shape Trios A/S due to a required software upgrade. This is a Class I recall affecting units distributed to hospitals, pharmacies, and laboratories in New Zealand. The recall requires affected devices to receive a software update to address the identified issue.

Recalling firm
Osstem New Zealand Limited
Product
TRIOS 6 System (L1P-2). [Device]. Model: TRIOS 6 intraoral scanner. Affected: 1ld2522l01055b, 1ld2535l01022b
Reason
The manufacturer has identified that a reflective surface inside the TRIOS 6 scanner front tube (frame 10032019) can lead to infrared (IR), straylight, which may produce image artefacts when using the SaMD TRIOS Dx Plus (proximal caries visualisation)., Artefacts may appear across multiple teeth in severe cases, or as localised glare in moderate cases. Localised glare may increase the likelihood of false positives for proximal caries, however, the risk of this problem leading to patient harm is low/negligible, as TRIOS Dx Plus is to be used only as an adjunct/aid. Direct visual and tactile...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-30
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/40808c73-cc87-4411-9e63-319d506205fc