Pride Mobility Products Corp.: Wheelchair, Powered

Recall Summary Pride Mobility is recalling Quantum Rehab Q6 Edge HD powered wheelchairs with software version 2.0.11 due to a controller defect. If users select the "Remind Me Later" option for reminders, pop-up messages may appear during wheelchair operation and cause the wheelchair to make an unexpected emergency stop or crash its control system, potentially resulting in falls or injury. The recall affects wheelchairs distributed nationwide in the US, as well as in Canada and Australia.

Recalling firm
Pride Mobility Products Corp.
Product
Brand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Reason
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-25
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/3f14bd01-c107-4655-b655-a517e65a8be1