Hamilton Medical Ag: Set, Tubing And Support, Ventilator (w Harness)

Hamilton Medical AG is recalling breathing circuit sets used with certain ventilator models (HAMILTON-C1, HAMILTON-T1, and HAMILTON-MR1) because the expiratory valve membranes may stick to the valve body in a way that cannot be detected before use, potentially causing an obstruction alarm and inadequate ventilation. The recall affects ventilator breathing circuit sets with specific reference numbers distributed nationwide across all U.S. states and territories. This is a Class I recall, the most serious classification.

Recalling firm
Hamilton Medical AG
Product
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Reason
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-29
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.

Official recall notice from FDA-DEVICES