Vanta Clinician Programmer Application - A71200 Clinician Programmer App

Recalling firm
Medtronic New Zealand Limited
Product
Vanta Clinician Programmer Application - A71200 Clinician Programmer App. [Device]. Model: Model: A71200 - Vanta Clinician Programmer Application; Model: 977006 - Vanta™ INS. Also covers: Vanta INS with AdaptiveStim™ Neurostimulator System
Reason
The manufacturer has released software ver 2.1.4148 of the Vanta Clinical Programmer Application (CPA), Model A71200. This software version and later versions, includes a firmware update to mitigate the previous communicated Action (Medsafe Ref#29701 and #31258)., Cardioversion procedures may damage the electronics of the Vanta™ Model 977006 implantable neurostimulator (INS), potentially resulting in loss of device function and requiring surgical replacement. To mitigate this risk, Medtronic has updated device labeling with recommended programming settings and considerations for...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-03
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE