Vanta Clinician Programmer Application - A71200 Clinician Programmer App
- Recalling firm
- Medtronic New Zealand Limited
- Product
- Vanta Clinician Programmer Application - A71200 Clinician Programmer App. [Device]. Model: Model: A71200 - Vanta Clinician Programmer Application; Model: 977006 - Vanta™ INS. Also covers: Vanta INS with AdaptiveStim™ Neurostimulator System
- Reason
- The manufacturer has released software ver 2.1.4148 of the Vanta Clinical Programmer Application (CPA), Model A71200. This software version and later versions, includes a firmware update to mitigate the previous communicated Action (Medsafe Ref#29701 and #31258)., Cardioversion procedures may damage the electronics of the Vanta™ Model 977006 implantable neurostimulator (INS), potentially resulting in loss of device function and requiring surgical replacement. To mitigate this risk, Medtronic has updated device labeling with recommended programming settings and considerations for...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-11-03
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE