Syringe Pumps_bd Swissco

Recalling firm
Becton Dickinson Ltd
Product
Syringe Pumps_BD SwissCo. [Device]. Model: BD Alaris™ neXus and Alaris™ Plus syringe pumps. Affected: Catalog number - 8002TIG03, 8002TIG03-G, 8003TIG03, 8003TIG03-G, 8005TIG03, CCneXus1, CCneXus1-S, PKneXus1 - All serial numbers
Reason
The manufacturer has identified through an internal investigation that the BD Alaris™ neXus and Alaris™ Plus syringe pumps may develop a mechanical anomaly on the linear potentiometer if the device is not stored in the correct position. The syringe plunger moves along the linear potentiometer component during infusion., This is an internal component and not readily visible to the Clinical User. If this anomaly occurs, it can lead to incorrect sensor readings, which may trigger error codes PL3, PL8, or PL9., These PL error codes are an intentional design feature that notify the Clinical...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-12
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE