Baxter Healthcare Corporation: Table, Operating-room, Electrical

Recalling firm
Baxter Healthcare Corporation
Product
Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
Reason
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among...
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-09-29
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution.

Official recall notice from FDA-DEVICES