[mdr] Philips Lumify Topical Transducer
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- [MDR] Philips Lumify Topical Transducer. [Device]. Model: Lumify Diagnostic Ultrasound System; Lumify Diagnostic Ultrasound System. Affected: All Software is impacted: Android Software versions: 1.X, 3.X, 4.X, 5.X iOS Software Versions: 2.X and 5.X; All Software is impacted: Android Software versions: 1.X, 3.X, 4.X, 5.X iOS Software Versions: 2.X and 5.X. Also covers: [MDR]Philips Lumify Ultrasound Systems
- Reason
- The manufacturer has identified the following potential problems with the Lumify Diagnostic Ultrasound System which may prevent or delay scanning functionality:, 1. Due to a software problem, the Lumify application may incorrectly require a previously registered transducer to be re-registered before use. Scanning cannot proceed until registration is completed, which requires Wi-Fi or cellular connectivity. In environments without network access, such as ambulances, aircraft, or remote settings, this may delay patient care., 2. The connection between the Lumify transducer USB cable and the...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-10-19
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/379fe338-a848-450f-843c-4e48c662640c