Medtronic Minimed, Inc.: Alternate Controller Enabled Insulin Infusion Pump

Recalling firm
Medtronic MiniMed, Inc.
Product
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
Reason
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-07-13
Status
Open, Classified
Distribution
International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/37599cfa-5a03-44cd-9a32-d88a98c58a27