Beckman Coulter Inc.: Analyzer, Chemistry (photometric, Discrete), For Clinical Use

Recall Summary Beckman Coulter is recalling DxC 500 AU Clinical Analyzer modules with software version 1.5 or lower due to a software defect that can erase customized test settings when new parameter packages are imported. When this happens, the analyzer may produce incorrect patient test results without alerting users to the problem. The recall affects clinical laboratories nationwide that use these analyzers.

Recalling firm
Beckman Coulter Inc.
Product
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers
Reason
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard;...
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-07-15
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide.