Hologic Inc: Selenia Dimensions Mammography Systems; System Name (model/catalog Number): 1. Selenia Dimensions 2d System Avia 3000/ (sda-sys-3000-2d-htc), 2. Se...
- Recalling firm
- Hologic Inc
- Product
- Selenia Dimensions Mammography Systems; System Name (Model/Catalog Number): 1. Selenia Dimensions 2D System Avia 3000/ (SDA-SYS-3000-2D-HTC), 2. Selenia Dimensions 2D System Avia 3000/ (SDA-SYS-3000-3D), 3. Selenia Dimensions 2D System 6000/ (SDM-SYS-6000-2D), 4. Selenia Dimensions 3D System 6000/ (SDM-SYS-6000-3D), 5. Selenia Dimensions Mammography System, Full Field Digital Mammography System, 2D (SDM-00001-2D), 6. Selenia Dimensions Mammography System 3D (SDM-00001-3D), 7. Selenia Dimensions Mammography System (SDM-00001-M2D), 8. Selenia Dimensions Mammography System (SDM-00001-M3D), 9....
- Reason
- The potential risk of sudden, unexpected freefall due to significant wear on the vertical lift nut of the C-Arm assembly on affected systems.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-11-02
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide TBD.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/35d082db-59c4-46f6-9cbe-591eb9ae66ee