Howmedica Osteonics Corp.: Tray, Surgical, Instrument
Recall Summary Howmedica Osteonics Corp. is recalling TS Misc. Instruments Trays (Triathlon, Catalog Number 6543-6-519) because the sterilization instructions may result in residual moisture remaining on the trays after sterilization when blue wrap is used. The trays were distributed to hospitals and medical facilities in the U.S. and approximately 158 units internationally through distributors.
- Recalling firm
- Howmedica Osteonics Corp.
- Product
- TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.
- Reason
- Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- Classification
- Class I
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-09-16
- Status
- Open, Classified
- Distribution
- These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.