Philips Medical Systems Nederland B.v.: Philips Azurion 7 M12. Model Numbers: 722078, 722223
Recall Summary Philips is recalling Azurion 7 M12 medical imaging systems (model numbers 722078 and 722223) worldwide because some units contain hard drives that weren't approved for use in these devices and may fail over time. If the drives fail, the system could lose imaging capabilities, show a blank screen, or be unable to save or access images. This is a Class II recall affecting devices distributed across the US and more than 100 other countries.
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Philips Azurion 7 M12. Model Numbers: 722078, 722223.
- Reason
- Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- U.S. Food and Drug Administration
- Recall date
- 2026-10-06
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands,...