Philips Medical Systems Nederland B.v.: Philips Azurion 7 M12. Model Numbers: 722078, 722223

Recall Summary Philips is recalling Azurion 7 M12 medical imaging systems (model numbers 722078 and 722223) worldwide because some units contain hard drives that weren't approved for use in these devices and may fail over time. If the drives fail, the system could lose imaging capabilities, show a blank screen, or be unable to save or access images. This is a Class II recall affecting devices distributed across the US and more than 100 other countries.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Azurion 7 M12. Model Numbers: 722078, 722223.
Reason
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-10-06
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands,...